7-4 MS-2 CCD Reader User’s Manual
Firmware
Firmware
Firmware Update
Application code versions are specific to your MS-2. Consult with your sales representative
before downloading application code. If needed, application code will be sent to you in the
form of a *.mhx file.
Contact Microscan’s help desk if you have difficulty downloading application code to the
Downloading Application Code to the MS-2
1. Connect to the MS-2 via ESP.
2. Verify your current firmware using Firmware Verification in ESP Utilities.
3. Power-off the reader by detaching the reader cable from the IC-3USB adapter (if you
are connected by USB) or IC-332 adapter (if you are connected by RS-232).
4.
Attach the plug end of the Firmware Upgrade Adapter (61-000143-01) to the reader cable.
5. Attach the socket end of the Firmware Upgrade Adapter to the IC-3USB adapter (if
you are connected by USB) or IC-332 adapter (if you are connected by RS-232) and
power-on the reader.
6. Click the Start button in ESP Utilities to open the firmware update utility.
7. Select the reader from the Target Device dropdown menu.
8. Select 8MHz from the Crystal Frequency dropdown menu.
(Continued on next page)
Firmware Upgrade Adapter
(61-000143-01)
Plug End Socket End
Click the Start button in
ESP Utilities to open the
firmware update utility.