Additional Notes about Calibration
Additional Notes about Calibration
The following conditions apply to the imager’s calibration process. Some of these items
are noted at various points throughout this section, or in other sections of the imager’s
documentation.
1. The Mirrored Image parameter <K514> is not part of the calibration
process.
2. Background Color <K451> is calibrated.
3. If Window of Interest Framing is enabled, the WOI will be set to full frame when
calibration begins. If WOI Framing is disabled, the current WOI configuration will be
used for the Search Pass.
4. If Symbol Type is enabled for calibration (Autodiscriminate), then Interleaved 2 of
5
Range Mode Status (Interleaved 2 of 5)
<K472>
will be enabled. This allows variable-length
Interleaved 2 of 5 symbols to be decoded. If an Interleaved 2 of 5 symbol is decoded
during calibration, then code length # 1 will be set to the decoded symbol length at the
end of calibration. Otherwise, the symbol lengths will be restored to their original
configurations.
5. Pharmacode is not calibrated.
6. All symbol types that were enabled before calibration will still be enabled after calibration.
For example, if Data Matrix ECC 200 was enabled before calibration, and the calibration
routine was performed on a Code 128 symbol, then after calibration is completed both
Data Matrix ECC 200 and Code 128 will be enabled.
7. The calibration process does not modify the global Composite status <K453>. The
global Composite status must be configured properly before calibration.
8. If the user requires calibration of a stacked symbology <K482>, <K483>, or <K484>
,
tha
t symbology must be enabled and configured appropriately before calibration.
9. UPC Supplementals will be disabled during calibration.
10. The Search process will use the configurable Window of Interest for image captures
.
However
, when the search process is completed, the WOI will be reduced
to include
o
nly the symbol of interest and some additional boundary area.
Vista de página 95
1 2 ... 91 92 93 94 95 96 97 98 99 100 101 ... 373 374

Comentários a estes Manuais

Vikitpm 04 Jan 2024 | 14:20:38

urenrjrjkvnm

Evapsi 06 Jan 2024 | 23:41:38

Med