Chapter 6 Autocalibration
MS-880 Industrial Long Range Scanner User’s Manual 6-9
6–Autocalibration
Note: Focus positions Inside, Outside and Set posted near the bottom of the above
menu are explained more fully in “Calibration Results Embedded Menu” on page 6-11.
Note: If you are using the 9-pin configuration port and doing autocalibration by serial
command or embedded menus, you will be able to view scanned bar code data in the
menu, but upon exiting the autocalibration menu, you will not be able to see bar code
data without sending the following command: <K146,,1>. If using the ESP, this is
not necessary.
Scanner Settings
Motor Speed, Gain Setting, and Tracking adjustments affect the read perfor-
mance. Focus controls a stepper motor that physically slides a lens back and forth to
control focal length. These settings are adjusted automatically in autocalibration.
The status of these parameters is displayed on the bottom line of the menu under
Motor, Gain, Track, and Focus (see “Overview of Optoelectrics” on page 8-4. These
settings are incremented or decremented by pressing the indicated key shown on the
menu. PDF-417 Decode
, AGC Status, Laser Power, and Raster Status use single
commands to toggle between states.
AGC Status
Typing S toggles AGC between Enable/Disable, as displayed on a the status line
near the bottom of the menu. With AGC disabled this value is simply the Gain setting.
However, with AGC enabled, this value represents the gain the scanner has to set to
maintain the desired signal amplitude. If a label is in close, the AGC level should be a
small value because the scanner does not need much gain to achieve the desired sig-
nal amplitude. However, as a label gets farther away from the scanner, this value will
increase to maintain the signal amplitude.
Laser Power
Typing L toggles between High and Low laser settings. See “Laser Power” on page 9-
8 for details.
PDF417 Decode Test
Typing D toggles the PDF417 Decode Test between Enable/Disable, as displayed
on a the status line near the bottom of the menu. Once set to Enabled, PDF417 only
is enabled and a special test mode is enabled that displays the decodes per second of
a full PDF417 label. While autocalibration is not performed in this mode, you can still
adjust settings manually.
This special read rate will remain enabled until the Autocalibration menu is exited or
PDF417 is disabled with another D command.
Note: This mode averages decode and capture time over a series of readings, thus
assuring for a smooth read-rate output, but at the cost of delayed reaction to label
movement and user setting changes.