Calibration Database by Embedded Menu
7-6 MS-880 Industrial Long Range Scanner User’s Manual
7–Calibration Data-
Calibration Database by Embedded Menu
To acc ess the Configuration Database menu from within the Autocalibration
Setup menu (page 6-8), type the number 3.
You can save your calibration settings in up to 7 database indexes.
1. Type D to save your present settings to the configuration database.
Note: From within the Autocalibration Setup menu, type the letter C.
The current Gain, Tracking, and Focus settings will be saved in Index 1.
Caution: Subsequent saves will overwrite your original settings unless you
change the # Active Settings to a number at least one greater than the num-
ber of index settings currently filled.
2. Change the # Active Settings to the number of index settings you expect to fill.
3. Press CR to return to the Autocalibration Setup menu and calculate the next
label position (if applicable).
4. Type D to save to the next index of autocalibration settings.
The database index will automatically increment by 1 and place your settings in
the next available index.
Note: See “Focus Position” on page 10-5 for additional ways for saving and
recalling database indexes.
5. Press the enter key (CR) to return to the main calibration menu or Esc to leave
the Autocalibration Setup menu entirely.